Bird - 4027F SeriesPrecision Filtered RF Power Sensors Filtered Power Measurement for Critical RF Applications The Bird 4027F Series RF Power Sensors are designed specifically for use in semiconductor processing and other precision process applications where the effects of amplitude modulation and harmonics need to be eliminated from the measurement. Because accuracy is critical, these sensors are capable of ±1% accuracy at the calibrated frequency and power level. Calibration is traceable to the National Institute of Standards and Technology (NIST), providing additional confidence in sensor measurement.

Bird4027F

PRODUCT FEATURES

  • Provides filtered measurement where the effects of amplitude modulation and harmonics need to be eliminated
  • ±1% accuracy at specified frequencies and power levels minimizing poor production yields
  • Models do not need to be field calibrated before use and only need to be calibrated once every six months
  • Plug and Play with the 4421A Series Multifunction Power Meter
  • Dozens of connector options available
  • Calibration traceable to the National Institute of Standards and Technology

FOR USE WITH

  • 4421A Multifunction Power Meters
  • SCC7 Calibration Cart
  • MSCC7 Calibration Cart

MODEL SELECTIONS

Model Freq. Range Power Range Cal. Freq. (Typical)
4027F2M 1.8 to 2.2 MHz 100 W to 10 kW 1.8, 2.0, 2.17 MHz
4027F10M 12 to 15 MHz 100 W to 10 kW 12.0, 12.5, 13.56, 14.0, 15.0 MHz
4027F60M 57 to 63 MHz 100 W to 6 kW 57.0, 58.5, 60.0, 61.5, 63 MHz
 

Here is the product specification described in a table in English:

Attribute Specification/Description
Item No. 4027F Series
Item Name Filtered Precision Directional Power Sensors
Frequency Range 1.8 MHz to 63 MHz (Sensor Dependent)
Power Range 100 W to 10 kW (Sensor Dependent)
Calibration Frequencies (Typical) 4027F2M: 1.8, 2.0, 2.17 MHz 
4027F10M: 12.0, 12.5, 13.56, 14.0, 15.0 MHz 
4027F60M: 57.0, 58.5, 60.0, 61.5, 63 MHz
Min Harmonic Rejection 4027F2M: 26 dB@3.6 MHz to 3.8 MHz, 30 dB@>3.8 MHz 
4027F10M: 30 dB@>25 MHz 
4027F60M: 30 dB@>114.0 MHz
LF Rejection 4027F2M: Not specified 
4027F10M: 30 dB@<1 MHz 
4027F60M: 30 dB@<15.0 MHz
Max Error Induced by 10% AM 4027F2M: 0.2%@<5 kW (1%@5kW to 10 kW)
4027F10M: 0.2%@<5 kW (1%@5kW to 10 kW) 
4027F60M: 0.2%@<1.5 kW (1%@1.5kW to 3 kW)
Accuracy ±1% at calibration frequencies and power levels 
±2% at other frequencies and power levels 
 Add 2% to uncertainty outside 25 ± 10 °C
Uniformity 2% maximum unit to unit, at calibration frequency and power levels
Speed 2 readings per second
VSWR Range 1.00 to 2.00
Directivity 28 dB
Insertion Loss <0.05 dB
Calibration Power Level 1700 W
Maximum Power 10 kW units: 12 kW (max) 
6 kW units: 7.2 kW (max)
Connectors Customer Specified
Recommended Calibration Interval 6 months. Performance before and after data to be supplied for units
Power Supply External DC, 12 VDC, supplied from Bird 4421A Power Meter
Temperature Operating: 0 to 50 °C (32 to 122 °F) 
 Storage: -20 to 70 °C (-4 to 158 °F)
Humidity 95% max (non-condensing)
Altitude Up to 10,000 feet (3,048 m)
Size 5.2 in x 2.5 in x 3.25 in (137 mm x 64 mm x 83 mm)
Weight 1 lb. 13 oz. (0.8 kg)
Certifications EMC: Designed to carry CE mark
 