Jofra DLC series temperature sensor
DLC series temperature sensor with Jofra RTC series temperature calibrator.
DLC dynamic load compensation system is another high-end innovation in the field of temperature calibration. DLC is used to compensate for the negative effects of the heat transfer between the dry calibration instrument and the measured sensor on the calibration accuracy.
When the sensor is inserted into the dry calibration instrument, the difference of the sensor size (diameter and length) and the difference between the calibration temperature and the ambient temperature forms an axial gradient deviation in the casing. We have been well known for the active dual calibration of the temperature calibrator, and the DLC has been further developed to compensate for the inside of the casing. The DLC can be combined with the RTC calibrator to be the only dry oven capable of compensating for the thermal loss generated by the sensor load.
• control the temperature of the casing under the accuracy of a percentage of the time
• the DLC indicator can display the activity and status of the two zones
When the temperature consistency is obtained, the DLC indicator will be displayed.
• simultaneous calibration of multiple sensors to save time
Calibration of large diameter sensors without reducing accuracy
No longer need to calibrate the sensor thermal part
option
DLC155 T differential TC. Probe length: 177 mm (6.96 in). Temperature range: -60 to 155 DEG C. Diameter: 3 mm (0.12 in) (outer diameter). Depth: 140 mm (5.51 in). Media compatibility: AISI316. Connection: Redel plug with memory
DLC158 T differential TC. Probe length: 196 mm (7.72 in). Temperature range: -60 to 155 DEG C. Diameter: 3 mm (0.12 in) (outer diameter). Depth: 140 mm (5.51 in). Media compatibility: AISI316. Connection: Redel plug with memory
DLC159 T differential TC. Probe length: 196 mm (7.72 in). Temperature range: -100 to 155 DEG C. Diameter: 3 mm (0.12 in) (outer diameter). Depth: 140 mm (5.51 in). Media compatibility: AISI316. Connection: Redel plug with memory
DLC250 T differential TC. Probe length: 196 mm (7.72 in). Temperature range: 0 to 250 DEG C. Diameter: 3 mm (0.12 in) (outer diameter). Depth: 140 mm (5.51 in). Media compatibility: AISI316. Connection: Redel plug with memory
DLC700 T differential TC. Probe length: 230 mm (9.05 in). Temperature range: 0 to 700 DEG C. Diameter: 4 mm (0.16 in) (outer diameter). Depth: 200 mm (7.9 in). Media compatibility: AISI316. Connection: Redel plug with memory
RTC-156 -30 to 155 DEG C (-22 to 311 DEG F)
RTC-158 -22 (-8 to 155 DEG to 311 DEG F dry type / tank)
RTC-159 -100 to 155 DEG C (-148 to 311 DEG F) super cooler
RTC-250 28 to 250 C (82 to 482 DEG F dry type / tank)
RTC-700 33 to 700 DEG C (91 to 1292 DEG F)
parts
127312 DLC155 casing
124897 DLC158 casing
128477 DLC159 casing
127759 DLC250 casing
127149 DLC700 casing
127587 Pt100 reference probe with a diameter of 3 mm
STS200A970 Pt100 reference probe with a diameter of 4 mm, suitable for DLC700
DLC series temperature sensor:
DLC155, DLC158, DLC250, DLC700